VDO MG 2100 Specifications Page 435

  • Download
  • Add to my manuals
  • Print
  • Page
    / 453
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 434
Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 435
Appendix G: Specifications
Specifications
Environmental specifications
The Vevo 2100 Imaging System operating environment should be free of fumes,
dirt, and electrical interference.
Specification Value
Temperature 10º to 40º C (50º to 104º F)
Relative humidity 15% to 80% non-condensing
Altitude Up to 2000m
System dimensions
Dimension Value
Height (without monitor) 112 cm (44 in.)
Height (with monitor) 155 cm (61 in.)
Width 71 cm (28 in.)
Depth 101 cm (39.5 in.)
Weight 170kg (375 lb.)
Ensure that sufficient clearance is available around the system for adequate
airflow and cooling. Do not block the air vents or air filters.
Electrical specifications
VisualSonics manufactures systems that operate with AC line voltages of 100V,
120V, and 240V. The electrical configuration of the system is noted on the system
nameplate.
100V
~
, 50/60Hz, 5A
120V
~
, 50/60Hz, 5A
240V
~
, 50/60Hz, 2.5A
For optimal system performance, use a dedicated, interference-free, isolated,
grounded/earthed wall outlet.