114 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 18: Setting up to acquire physiological data
Auto-calibrating your Vevo-supported blood pressure instrument
The Vevo 2100 Imaging System includes pre-configured calibration settings for
the Millar PCU-2000 Pressure Control
f To calibrate a Vevo-supported blood pressure instrument:
1. Connect the pressure instrument to the Advanced Physiological Monitoring
Unit and ensure that the Advanced Physiological Monitoring Unit is
connected to the Vevo 2100 Imaging System at the Physio Data connector on
the rear panel of the system. Ensure that all three systems are powered on.
2. Open an image mode window by beginning to acquire data in any imaging
mode or opening any image from the Study Browser.
3. Press
Physio Settings
.
The left panel displays the physiological display setting sections.
4. In the Blood Pressure section:
a. In the upper drop-down list select Import Calibration.
b. In the lower drop-down list select the preconfiguration for your pressure
monitor.
c. Click Calibrate.
5. The system:
Calibrates your pressure scale.
Retains the calibration settings between imaging sessions. You only need
to repeat the calibration procedure if you connect a different blood
pressure monitor or if you think there might be a problem with the
calibration accuracy.
Manually calibrating any blood pressure instrument
The Vevo 2100 Imaging System can calibrate any blood pressure scale manually,
as long as it includes a built-in calibration function.
f To calibrate any blood pressure instrument:
1. Connect the pressure instrument to the Advanced Physiological Monitoring
Unit and ensure that the Advanced Physiological Monitoring Unit is
connected to the Vevo 2100 Imaging System at the Physio Data connector on
the rear panel of the system. Ensure that all three systems are powered on.
2. Open an image mode window by beginning to acquire data in any imaging
mode or opening any image from the Study Browser.
3. Press
Physio Settings
.