116 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 18: Setting up to acquire physiological data
gate period is called the window and is defined by a dark yellow background that
follows the trace across the screen.
Window
Delay
Respiration trace (yellow)
Before you begin:
Your animal must be connected to the Advanced Physiological Monitoring
Unit.
In the Physiological Enable section of the General tab in the Preferences
window, the Respiration check box must be selected
IMPORTANT: You can only activate and control respiration gating while you are
acquiring data. You cannot access these options when you review an image.
f To activate respiration gating:
1. Begin acquiring data.
2. Press
Physio Settings
.