Page view 194
Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 195
Chapter 33: Acquiring B-Mode images
Image Depth
Adjusts how deep in mm you want to display the ultrasound signal. Pull down to
increase the depth. Push up to decrease the depth. The available depth is transducer
dependent.
Focus Depth
Adjusts the depth of the B-Mode focal zone or focal zones on your image. When you
have more than one focal zone this control moves the depth of all the focal zones as a
group. Push up to decrease the depth. Pull down to increase.
Focal Zones
This control adjusts the number and configuration of focal zones on your B-Mode based
image.
Focal zones enhance the resolution across your image, while slightly reducing the
acquisition frame rate. The system always displays at least one focal zone, and you can
apply a maximum of two additional zones depending on the transducer. When you add
focal zones the system maximizes the resolution for a larger area of your image, and
reduces the acquisition frame rate.
To use this rocker switch control:
1. Push the rocker switch forward to cycle through the following focal zone application
sequence:
Single zone
Two zones, narrow
Two zone, wide
Three zones, narrow
Three zones, wide
2. Pull the rocker switch back to cycle back through the focal zone options in reverse.