Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 401
Appendix B: Troubleshooting
Troubleshooting
If a problem is encountered when using the Vevo 2100 Imaging System, try the
solutions described in this appendix. If none of the solutions solves the problem,
contact a VisualSonics Technical Support representative
System panel controls
Problem Solution
System does not power up
Ensure that the main power cable for the system is properly
connected to the Vevo 2100 Imaging System.
Ensure that the system is plugged into a grounded/earthed
wall outlet. Turn the main power switch On.
Turn the computer standby switch On.
No audio
Adjust the Volume dial
Adjust any PW Doppler settings (such as the PW Doppler
angle, the Doppler Gain, the Sample Volume Position) to
increase the strength of the PW Doppler signal.
Study Browser
Problem Solution
Unable to create new studies
Ensure that an transducer is connected to the front panel of the
Vevo 2100 Imaging System, and ensure that it has been initialized.
Unable to commit a study session Ensure that an operator has been specified.
The system tells you that your study
is corrupted
Cause: You are still in an active image acquisition session working
on an active series. The system cannot open the Study Info window
until you close the series. This prevents you from accidentally
leaving and closing the series before you have added all the
required images to your series.
Solution: Click Close Series now or complete all the images you
need to acquire for your series and then return to the Study Browser
and click Close Series. Then press
Study Info
.
Appendix B